Welcome: Shanghai cixi instrument CO.,LTD,it's a professional 35 years manufacturer of micro flow meter,ultrasonic flow meter,water flow meter,magnetic flow meter,turbine flow meter,vortex flow meter,coriolis flow meter,digital flow meter,thermal mass flow meter,liquid flow meter,gas flow meter,dosing pump,gear metering pump,level transmitter,level sensor etc.

OIL/GAS

Home > Application > OIL/GAS >

Three Thermal Mass Flow Meters for a NOx Test Rig Simpler Than Venturi + DP

Three Thermal Mass Flow Meters for a NOx Test Rig: Simpler Than Venturi + DP
 
When specifying flow measurement for a NOx generator test rig, the choice often gets stuck between two extremes: a Venturi tube plus differential pressure transmitter (low headline cost, but hidden engineering in piping modifications, temperature/pressure compensation and commissioning), or premium Coriolis/imported thermal meters (excellent performance, but three units blow the budget).
 
The customer's technical team proposed a more pragmatic approach: three thermal mass flow meters in different configurations, one per measurement zone, all with 316 stainless wetted parts, 24 VDC supply and 4–20 mA output. Here is a concise summary for similar applications.
 
Zone A — Combustion Air (DN100 Insertion Type)
 
Duty: Blower suction side, 110.3 mm ID, 80–140 m³/h at 20°C, near atmospheric (slightly negative), extremely low pressure loss required. Standard DN100 acceptable if priced better.
 
Recommendation: DN100 insertion thermal mass flow meter, 316 SS. Insertion design avoids cutting the pipe for a flanged spool, adds negligible pressure drop, and costs far less than inline meters at this diameter. Thermal technology outputs mass flow or normalized volume flow directly — no temperature/pressure compensation needed, just one clean 4–20 mA signal to the PLC.
 
Accuracy and pressure drop: Typical economy thermal meters deliver plus or minus 1–2 percent of reading; the insertion probe's obstruction to airflow is minimal, so pressure loss is effectively negligible — critical on a blower suction line.
 
Zone B — Heated Air to Gas Treatment Module (DN25 Inline, High-Temperature)
 
Duty: 21 mm ID (DN25), 4.2–7.0 m³/h at 20°C up to 6.8–11.3 m³/h at 200°C; normal max 200°C, absolute max 400°C; slight positive pressure (a few mbar). Sensor and probe must withstand 400°C; split design with separate electronics is acceptable.
 
Recommendation: DN25 inline high-temperature thermal mass flow meter, 316 SS, split design. The 400°C limit is the binding constraint — standard thermal sensors are typically rated below 200°C, so a high-temperature probe is mandatory. Split electronics keep the transmitter away from the hot pipe. Inline is preferable to insertion at DN25 for compactness and stability.
 
Specifying note: State "sensor probe rated to 400°C" explicitly in the RFQ and request material compatibility documentation. Some economy high-temp models reach 350–400°C at the probe — confirm it is the probe itself, not just a heat sink.
 
Zone C — Dutch Natural Gas (G-gas), DN15 Inline
 
Duty: 15 mm ID (DN15), 7.3–12.2 m³/h at 20°C atmospheric, supply pressure up to 200 mbar(g), methane-rich pipeline gas. Must be suitable for flammable gas, calibrated for G-gas composition, and low pressure drop.
 
Recommendation: DN15 inline thermal mass flow meter, 316 SS, flanged. Two hard requirements for natural gas: suitability for flammable service (Ex or intrinsically safe certification depending on the installation zone), and calibration on the actual gas composition — thermal meters rely on the gas's specific heat and thermal conductivity, so an air-calibrated unit used on natural gas will carry significant error. Provide the typical G-gas composition (usually more than 85 percent methane with minor nitrogen and CO₂) and require calibration to it.
 
Alternative considered: A DN25 vortex meter (316 SS, flanged, 400°C) was also mentioned. At DN25 on gas, vortex meters have weak signal at low flow and require temperature/pressure compensation to obtain mass flow — more system complexity than thermal. If the Zone C lower flow of 7.3 m³/h gives adequate velocity in DN15, thermal remains the simpler choice.
 
Common Requirements for the RFQ
 
Power: 24 VDC. Output: 4–20 mA proportional to mass flow or Nm³/h. Accuracy: plus or minus 1–2 percent of reading. Wetted parts: SS316/316L. Pressure drop: very low, negligible for insertion and to be confirmed for inline. Certification: CE with technical documentation. PLC interface: standard 4–20 mA input.
 
Indicative pricing: Economy thermal mass flow meters in DN15 and DN25 inline versions typically fall in the low-thousands RMB range per unit; the DN100 insertion unit is somewhat higher due to probe length and mounting hardware, but the three-unit total stays well within budget. Firm quotes require the duty parameters above.
 
Versus Venturi + DP
 
The customer was comparing a Venturi plus differential pressure transmitter system. To be fair, DP has an advantage in Zone B's 400°C service — the transmitter can sit away from the hot tap — but the Venturi itself must still withstand the temperature, and micro-differential measurement at a few mbar positive pressure has poor signal-to-noise. The three-thermal-meter approach wins on simplicity: each meter outputs mass flow independently, no temperature/pressure compensation math, and minimal PLC configuration effort. For a non-standard, small-batch test rig where commissioning time is precious, simplification is itself a value.
 
Next step: Send the RFQ to two or three domestic economy thermal meter suppliers in parallel, explicitly flagging Zone B's 400°C probe requirement and Zone C's G-gas composition calibration. 


If you also have similar requirements for precision flow meter or industrial gas flow meter selection and quotation, feel free to contact
CIXIFM Winny:
Email:
[email protected], WhatsApp: 86 18049841995





 

PRODUCTS

CATEGORIES

CONTACT US

Contact:Leon

Tel: +86021-57635022

Email:[email protected]

Add:Room 402, No. 650, Xinzhuan Rd., Jiuting Town, Songjiang District, Shanghai, China (Mainland)